SPECIFICATIONS
Model name | HV-50 II | HV-80 II | HV-110 II | |
Classification of MD/LB | MD | |||
Chamber dimension | Ø300 x D710 mm | Ø420 x D615 mm | Ø420 x D795 mm | |
Chamber volume | 54.2L | 98.8L | 123.7L | |
Effective capacity | 50L | 85L | 110L | |
Temp. settings | Strilization temperature | 105 - 135 °C Variable | ||
Dissolution temperature | 60 - 100 °C Variable | |||
Warming temperature | 45 - 60 °C Variable | |||
Lid lock temperature | LIQUID, AGAR, DISSOL modes: 60-95°C variable, SOLID, WASTE modes: 60-97°C variable | |||
Time settings | Strilization timer | 1-250 min, The remaining time is displayed. | ||
Dissolution temer | 1-60 min, The remaining time is displayed. | |||
Auto-start timer | 1 min later-1 week later, A start date and time is set | |||
Air removal time | 3-9 min variable6-18 min variable | |||
Mode selection | LIQUID, SOLID, WASTE, AGAR, DISSOL/8 programs | |||
Exhaust settings | 3 steps (No exhaust, Fine exhaust, Small amount of exhaust) | |||
Maximum working pressure | 0.255MPA | |||
Thermometer | Digital disaply 5.0 - 137.9°C | |||
Pressure guage | Digital display 0-0.3 MPa/Analog display 0-0.4 MPa | |||
Safety devices and alarms | Safe pressure valve, Earth leakage breaker with O.C,protection, Low water cut off device, Alarms: Low water heating, Temperature sensor wire disconnection, Over-temperature, Over-cool, Over-pressure, Abnormality in lid, Abnormality in the lock of lid opening/closing lever | |||
Sterilization timer | 1 - 250 Min, the remaining time is displayed | |||
Outside dimention (in mm) | W 547 x D 532 x H 1046 | W 667 x D 652 x H 1,026 | W 667 x D 652 x H 1,180 | |
Air removal time | 3 - 9 min Variable | 6 - 18 min Variable | ||
Weight (approx.) | 60 kg | 83 kg | 89 kg | |
Chamber material | Stainless steel (SUS 304) | |||
Power souree | AC 110/120/220/230/240, 50/60Hz | AC 220/230/240, 50/60Hz | ||
Power consumption | 2.0kW | 3.0kW | 4.0kW | |
Power cord connection | Ring terminal (A plug is not supplied.) |