SPECIFICATIONS
Temperature
Model | HCP50 | HCP105 | HCP150 | HCP240 |
Working-temperature range | +7 above ambient temperature up to +90 °C | |||
Setting temperature range | +18 to +90 °C | |||
Setting accuracy temperature | 0.1 °C | |||
Temperature sensor | 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error |
Humidity
Model | HCP50 | HCP105 | HCP150 | HCP240 |
Humidity Control | Active microprocessor control for humidifying and dehumidifying (20 95 % rh), incl. digital indication and auto-diagnostic system ensures even more rapid reaching of set humidity and very short recovery times. Humidity supply with water (only for demineralised water with a conductivity of 5 to 10 μS/cm and a pH value between 5 and 7; from an external tank) by a self-priming pump; integral bacteria block by generating hot steam, dehumidifying via sterile filter | |||
Setting range active humidity control | 20 to 95 % rh and rh-Off | |||
Setting accuracy humidity | 0.5 % rh |
Stainless steel interior
Model | HCP50 | HCP105 | HCP150 | HCP240 |
Dimensions (d less 35 mm for fan) | 400 x 425 x 330 mm | 560 x 480 x 400 mm | 560 x 700 x 400 mm | 600 x 810 x 500 mm |
Interior | W.-St. 1.4301 (ASTM 304) | |||
Volume | 56 l | 107 l | 156 l | 241 l |
Max. number of internals | 5 | 6 | 10 | 12 |
Max. loading of chamber | 75 kg | 90 kg | 120 kg | 140 kg |
Max. loading per internal | 15 kg |
Textured stainless steel casing
Model | HCP50 | HCP105 | HCP150 | HCP240 |
Dimensions (d +56mm door handle) | 559 x 795 x 521 mm | 719 x 850 x 591 mm | 719 x 1070 x 591 mm | 719 x 1070 x 591 mm |
Electrical data
Model | HCP50 | HCP105 | HCP150 | HCP240 |
Voltage Electrical load | 230 V, 50/60 Hz approx. 1520 W | 230 V, 50/60 Hz approx. 1720 W | 230 V, 50/60 Hz approx. 1800 W | 230 V, 50/60 Hz approx. 1840 W |
Voltage Electrical load | 115 V, 50/60 Hz approx. 1520 W | 115 V, 50/60 Hz approx. 1720 W | 115 V, 50/60 Hz approx. 1800 W | 115 V, 50/60 Hz approx. 1840 W |
Ambient conditions
Model | HCP50 | HCP105 | HCP150 | HCP240 |
Ambient temperature | 10 °C to 30 °C | |||
Set Up | The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm. | |||
Humidity rh | max. 70 %, non-condensing | |||
Altitude of installation | max. 2,000 m above sea level | |||
Overvoltage category | II | |||
Pollution degree |
ตู้ควบคุมอุณหภูมิและความชื้น
เครื่องลดความชื้น
ตู้ควบคุมอุณหภูมิความชื้น
ตู้ควบคุมอุณหภูมิและความชื้นคงที่
Humidity Chamber
Chamber
Humidity & Temp Chamber Memmert
Temperature and Humidity Test Chamber
HCP50
HCP105
HCP240
Memmert
HCP150
สินค้าที่เกี่ยวข้อง
เว็บไซต์นี้มีการใช้งานคุกกี้ เพื่อเพิ่มประสิทธิภาพและประสบการณ์ที่ดีในการใช้งานเว็บไซต์ของท่าน ท่านสามารถอ่านรายละเอียดเพิ่มเติมได้ที่
นโยบายความเป็นส่วนตัว และ นโยบายคุกกี้ เปรียบเทียบสินค้า
0/4 ลบทั้งหมด
เปรียบเทียบ
Powered By
MakeWebEasy
Home Home Home Home |