SPECIFICATIONS
MODEL | CM-5 |
Illumination/viewing system | Reflectance: di:8 °, de:8 ° (diffuse illumination: 8 ° viewing) SCI (specular component included)/SCE (specular component excluded) switchable Conforms to CIE No. 15, ISO 7724/1, ASTM E 1164, DIN 5033 Teil 7, and JIS Z 8722 (Condition c) Transmittance: di:0 °, de:0 ° (diffuse illumination: 0 ° viewing) |
Integrating sphere size | Φ152 mm |
Detector | Dual 40-element silicon photodiode arrays |
Spectral separation device | Planar diffraction grating |
Wavelength range | 360 nm to 740 nm |
Wavelength pitch | 10 nm |
Half bandwitdh | Approx. 10 nm |
Measurement range | 0 to 175 % (Reflectance or transmittance); Output/display resolution: 0.01% |
Light source | Pulsed xenon lamp |
Measurement/illumination area | Reflectance: Changeable by changing mask and settings. LAV:Φ30 mm/Φ36 mm; MAV (optional):Φ8 mm/Φ11 mm; SAV (optional):Φ3 mm/Φ6 mm Transmittance: Φ26 mm |
Repeatability | Spectral reflectance: Standard deviation within 0.1% (400 nm to 740 nm) Chromaticity value: Standard deviation within ΔE*ab 0.04 * When a white calibration plate is measured 30 x at 10-second intervals after white calibration |
Inter-instrument agreement | Within ΔE*ab 0.15 (Typical) (LAV/SCI) (Based on 12 BCRA Series II color tiles compared to values measured with a master body under Konica Minolta standard conditions) |
Transmittance chamber | No sides (unlimited sample length); Depth (maximum sample thickness): 60 mm Sample holders (optional) for holding sheet samples or containers of liquid samples can be installed/removed |
Display | 5.7-inch TFT color LCD |
Display languages | English, Japanese, German, French, Italian, Spanish, Simplified Chinese, Portuguese |
Operating temperature/humidity range | 13 to 33 °C, relative humidity 80% or less with no condensation |
White/100% calibration | Automatic white (reflectance)/100% (transmittance) calibration using internal white calibration plate (Not applicable to 100% calibration when using cells for transmittance measurements of liquids.) |
Interfaces | USB 1.1 (Connection to PC; USB memory stick); RS-232C standard (Connection to serial printer) |
Observer | 2 ° Standard Observer or 10 ° Standard Observer |
Illuminant | A, C, D50, D65, F2, F6, F7, F8, F10, F11, F12, ID50, ID65 (simultaneous evaluation with two light sources possible) |
Displayed data | Spectral values/graph (reflectance, transmittance, absorbance), colorimetric values/graph, color-difference values/graph, pass/fail judgment, pseudo color, color assessment |
Color space | L*a*b*, L*C*h, Hunter Lab, Yxy, XYZ, Munsell, and color differences in these spaces (except for Munsell) |
Index | Reflectance: MI; WI (ASTM E 313-73, ASTM E 313-96); YI (ASTM E 313-73, ASTM E 313-96, ASTM D 1925); ISO Brightness; B (ASTM E 313-73) Transmittance: Gardner; Iodine Color Number; Hazen/APHA; European Pharmacopoeia; US Pharmacopeia |
User index | User-defined index |
Color-difference equation | ΔE*ab (CIE 1976), ΔE*94 (CIE 1994), ΔE00 (CIE 2000), ΔE (Hunter), CMC (l: c) |
Pass/fail judgment | Tolerances can be set to colorimetric values (except Munsell), color-difference values, or reflectance index values |
Storable data | Measurement data: 4,000 measurements; Target color data: 1,000 measurements |
USB memory stick* storage | Storage of measurement data and target color data. Storage/reading of measurement condition settings * Security-enabled USB memory sticks cannot be used. |
Power | AC 100 to 240 V, 50/60 Hz (using exclusive AC adapter) |
Size | Slide cover closed: 385 (W) x; 192 (H) x; 261 (D) mm Slide cover open: 475 (W) x; 192 (H) x; 261 (D) mm |
Weight | Approx. 5.8 kg |
Operating temperature/humidity range | 13 to 33 °C, relative humidity 80 % or less (at 33 °C) with no condensation |
Storage temperature/ humidity range | 0 to 40 °C, relative humidity 80 % or less (at 33 °C) with no condensation |